DC Characteristics
|
|
Function
|
Range
|
Resolution
|
Input Resistance
|
1 Year Accuracy (% of reading
+% of range) (23C°±5C°)
|
|
DC Voltage[3]
|
100.0000mV
|
0.1µV
|
>10G Ohm
|
0.0050+0.0035
|
|
1.000000V
|
1.0µV
|
>10G Ohm
|
0.0040+0.0007
|
|
10.00000V
|
10µV
|
>10G Ohm
|
0.0035+0.0005
|
|
100.0000V
|
100µV
|
10M Ohm
|
0.0045+0.0006
|
|
1000.000V
|
1mV
|
10M Ohm
|
0.0045+0.0010
|
|
Function
|
Range
|
Resolution
|
Shunt Resistance
|
1 Year Accuracy (% of reading
+% of range) (23C°±5C°)
|
|
DCI (DC Current)
|
10.000000mA
|
10 nA
|
5.1 Ohm
|
0.05+0.020
|
|
100.00000mA
|
100 nA
|
5.1 Ohm
|
0.05+0.005
|
|
1.0000000A
|
1 µA
|
0.1 Ohm
|
0.100+0.010
|
|
3.0000000A
|
10 µA
|
0.1 Ohm
|
0.120+0.020
|
|
Function
|
Range
|
Resolution
|
Test Current
|
1 Year Accuracy (% of reading
+% of range) (23C°±5C°)
|
|
Resistance (4)
|
100.0000 Ohm
|
100 µOhm
|
1mA
|
0.010+0.004
|
|
1.000000 KOhm
|
1 mOhm
|
1mA
|
0.010+0.001
|
|
10.00000 KOhm
|
10 mOhm
|
100 µA
|
0.010+0.001
|
|
100.0000 KOhm
|
100 mOhm
|
10 µA
|
0.010+0.001
|
|
1.000000 MOhm
|
1 Ohm
|
5 µA
|
0.010+0.001
|
|
10.00000 MOhm
|
10 Ohm
|
500 nA
|
0.040+0.001
|
|
100.0000 MOhm
|
100 Ohm
|
500 nA//10MOhm
|
0.800+0.010
|
|
|
AC Characteristics (7)
|
|
Function
|
Range(8)
|
Resolution
|
Frequency (Hz)
|
1 Year Accuracy (% of reading
+% of range) (23C°±5C°)
|
|
ACV (AC TRMS Voltage) (9)
|
1000.0000mV
|
0.1µV
|
3-5
|
1.00+0.04
|
|
5-10
|
0.35+0.04
|
|
10-20K
|
0.06+0.04
|
|
20-50K
|
0.12+0.05
|
|
50K-100K
|
0.60+0.08
|
|
100K-300K
|
4.00+0.50
|
|
1.000000V To 750.000V
|
1.0 µV To 1mV
|
3-5
|
1.00+0.03
|
|
5-10
|
0.35+0.03
|
|
10-20K
|
0.06+0.03
|
|
20-50K
|
0.12+0.05
|
|
50K-100K
|
0.60+0.08
|
|
100K-300K
|
4.00+0.50
|
|
ACI (AC TRMS Current) [11]
|
1.000000A
|
1 µV
|
3-5
|
1.00+0.04
|
|
5-10
|
0.30+0.04
|
|
10-5K
|
010.+0.06
|
|
3.000000A
|
10 µV
|
3-5
|
1.10+0.06
|
|
5-10
|
0.35+0.06
|
|
10-5K
|
015+0.06
|
|
Diode Test
|
1.0000V
|
10 µV
|
1 mA
|
0.010+0.020
|
|
Continuity
|
10000.00kOhm
|
10 mOhm
|
1 mA
|
0.0100.030
|
|
Frequency and Period
|
|
Function
|
Range
|
Frequency
|
1 Year Accuracy (% of reading
+% of range) (23C°±5C°)
|
|
Frequency & Period
|
100mV to 750V
|
3-5
|
0.10
|
|
5-10
|
0.05
|
|
10-40
|
0.03
|
|
40-300K
|
0.01
|
|