PTS
Portable DC Test Set Series
|
|
The PTS Portable DC Test Set are combination high voltage
DC proof testers, or hipots, and high voltage Megohmmeters.
Testing for dielectric strength and insulation resistance
is now served with one instrument, saving money, size and
weight. The PTS series is designed and constructed with field
use in mind. It is rugged, reliable and contains all of the
features needed for electrical maintenance testing of new
or installed electrical apparatus.
|
PTS-75 Model (0-75 kVdc, 10mA)
|
|
Advantage
|
- Five range current meter with a 0-1.0
µA low range for Leakage Current measurements with
resolution to 10 nanoamperes
- Megohm readings at any output voltage
- Regulated input for accurate, reliable
results. Ideal for field generator use. ± 1% stability with
± 10% line change (PTS-37.5 & PTS-75 models)
- Internal HV shorting solenoid with
discharge resistor
- Continuous Duty Cycle
- Automatic transit protected meters
- Rugged case with cushion grip handles
- Glass front meters eliminate static
buildup
|
| Application |

PTS-130 Model (0-130 kVdc, 10mA)
|
| The PTS Series can be
used as a high voltage DC Proof/Hipot tester or as a high voltage
Megohmmeter. It is used to proof test or megohm cable insulation,
switchgear, motors, transformers, insulators and other electrical
apparatus. Its combination Hipot/HV Megohmmeter capability makes
it extremely versatile, especially for field testing applications. |
| Features |
- Two range voltmeter and five range current meter with
a 0-1.0 µA low range with
resolution to 10 nanoamperes
- Output current limited to 10 mA, 5 mA on PTS-200
- Internal HV shorting solenoid with discharge resistor
(PTS-37.5 through PTS-130)
- Internal discharge/bleed down resistors on PTS-200
- Automatic transit protected meters
- Insulated Return and Guard circuit to bypass stray and
unwanted leakage current
- Rugged case with storage for all input/output leads and
ground stick (PTS-37.5 through PTS-80)
- Cable storage bin on HV tank (PTS-100 through PTS-200)
- Surge and transient protection
- External safety and zero start interlocks
- Glass front meters eliminate static buildup .
|
|
PTS Model Specifications
|
|
|
PTS-37.5
|
PTS-75
|
PTS-80
|
PTS-100
|
PTS-130
|
PTS-200
|
|
Input
|
60Hz
|
120V, 5 Amps
|
120V,
10 Amps |
120V,
10 Amps |
120V,
10 Amps |
120V,
15 Amps |
120V,
15 Amps |
|
50/60Hz
|
230V, 3 Amps
|
230V,
5Amps |
230V,
5Amps |
230V,
5Amps |
230V,
8Amps |
230V,
8Amps |
|
Output
|
0-37.5kVdc, 10mA
|
0-75kVdc,
10mA |
0-80kVdc,
10mA |
0-100kVdc,
10mA |
0-130kVdc,
10mA |
0-200kVdc,
10mA |
|
Polarity
|
Negative Output, Positive Ground
|
|
Duty
|
Coutinuous, Capacitive Charging
|
|
Ripple
|
N/A Capacitive Load. Less Thank 2.5% RMS
Resistive Load
|
|
Regulation
|
<1% with 10% input line change capacitive
load
|
N/A (no ferro-resonant regulator)
|
|
Kilovoltmeter
|
0-15/37.5 kVdc
|
0-37.5/75
kVdc |
0-40/80
kVdc |
0-50/100
kVdc |
0-75/130
kVdc |
0-80/200
kVdc |
|
Current Meter (Megohm Scale)
|
0-1.0 dc microamperes, w/x1, x10, x100,
x1k, x10k, 100-1 Megohms, w/0.1, x10, x100, x1k
|
|
Control Dimensions
|
356x279x356(mm)
|
356x279x457(mm) |
356x279x457(mm) |
241x298x368(mm) |
318x306x457(mm) |
330x368x775(mm) |
|
H.V. Tank Dimension
|
Included in Control
|
Included
in Control |
Included
in Control |
241x298x368(mm) |
318x305x457(mm) |
330x368x775(mm) |
|
Control Weight
|
50 lbs (23kg)
|
68
lbs (31kg) |
62
lbs (28kg) |
30
lbs (14kg) |
34
lbs (15kg) |
34
lbs (15kg) |
|
Tank Weight (With Cable)
|
|
|
|
68
lbs (31kg) |
82
lbs (37kg) |
150
lbs (68kg) |
|
Output Termination
|
Shieled output cable with clip and boot,
20 ft (6m)
|
|
Additional Cable and Accessories
|
Input line cord, 10 ft, Return lead with
clip and boot, 20 ft.(6M), Ground lead with clip and boot,
20 ft. (6M), Ground stick with 20 ft. (6M) lead and clip
|
Back
to the instrument selecion
For your price and delivery inquires, application
details comments, and general feedback,
please select Feedback
Mailbox
|