| Specification |
| Measurement
Parameters |
Voltage/Turns Ratio Range:
|
0.8 ... 13,000 |
Resolution:
|
5 Digits |
| Accuracy |
Range 0.80 ...2000
|
±0.08% with 100 and 40 Volt, ±0.15%
with 10 Volt |
Range 2001 ...4000
|
±0.15% with 100 and 40 Volt, ±0.22%
with 10 Volt |
Range 4001...13,000
|
±0.3% with 100 and 40 Volt, ±0.5%
with 10 Volt |
| Current Measurement |
Range :
|
0 ... 1 Ampere |
Accuracy:
|
± 0.001 Ampere |
Resolution:
|
0.0001 Ampere |
| Phase angle |
Range :
|
± 90 Degrees |
Accuracy:
|
± 0.05 Degree |
Resolution:
|
0.01 Degree |
| Size
|
L: 18. 5" (470mm)
W: 14.6" (371mm) H: 7.5" (190mm) |
| Weight |
18 lbs. (8.2 kg). |
| Input
Power |
100 to 250 vac
50/60 Hz Autoranging. Fuse: 2 Amp |
| Test
Voltage |
User selectable:
100, 40, 10 and 1 vac 1 ampere |
| Panel
Display |
LCD graphic with
back lighting |
| Front
Panel |
Sealed, Piezoelectric
actuation |
| Interface |
9 pin RS232 serial
/ 25 pin standard centronics parallel |
| Memory
Storage |
Internally stores
more than 1,300 locations in memory (100 complete tests) |
| Temperature |
Operating - 10°
C to 60° C Storage - 40° C to 70° C |
| Cable
Set / Accessories |
3-phase
cable set: length, total length 44 feet ( 12 feet long with
an extension cable of 32 feet long ) Cable carrying bag, power
supply cord, instruction manual |